AfiMilk MPC Installation Manual
79
First Time Operation
Chapter
7
Statistical Accuracy Calculation
After scaling milk weights for all the milk meters in the parlor, statistical accuracy
calculation is performed for each milk meter.
Statistical calculation is done for each reading and corresponding weigh scale
measurement for calculating the difference as a percentage value between the reading of
the milk meter and the actual milk weighed.
The following formula is used for calculating difference between the milk meter reading
and the milk weighed.
Average all of the DIF% deviations for each Milk Meter.
The average deviation for every milk meter should be less than 3% to be considered as an
accurate measuring device.
Use the G & H adjustment procedure described above for adjusting inaccuracies in
measurements of specific milk meters.