AfiMilk MPC Installation Manual 76
Chapter
7
First Time Operation
Adjusting the “G” Parameter Value
Change the value of "
G
" parameter per calculated average deviation. If the result is
negative (Milk meter reading is lower than the actual weight), the calculated adjustment
value needs to be added to the value of
G
. If the adjustment is positive, the calculated
value needs to be subtracted from the value of
G
.
General adjustment calculation →%Dev X (–2.5) = Adjustment value
Example: if the actual value of G in the milk meter is 55 and the measured deviation -4%
(Milk meters show 4% less than actual weight), modify the “G” parameter
value by:
-4 X (-2.5) = +10 to 65
Use the same procedure explained above for Air Admission Compensation for accessing
and changing the value of G parameter.
Adjusting the “H” Parameter Value
Change the value of "
H
" parameter per calculated average deviation. If the result is
negative (Milk meter reading is lower than the actual weight), the calculated adjustment
value needs to be added to the value of
H
. If the adjustment is positive, the calculated
value needs to be subtracted from the value of
H
.
General adjustment calculation →%Dev X (–5) = Adjustment value
Example: if the actual value of H in the milk meter is 37 and the measured deviation -4%
(Milk meters show 4% less than actual weight), modify the “H” parameter
value by:
-4 X (-5) = +20 to 57
Check the accuracy of the new settings within a few days after setting the
new “G” value. This may be done by comparing total milk collected by
the afimilk system to the bulk tank accumulation.