Image formats DICOM (Default)
TIFF
Postscript Not available
Power consumption - heat dissipation
During operation 250 W - 900 kJ/h
In stand-by 70 W - 252 kJ/h
Peak power (absolute max. rating) 530 W - 1908 kJ/h
Protection against
Electrical shocks Class 1 (grounded)
Ingress of water IPXØ
Environmental conditions (operation)
Room temperature Between +15°C and +35°C
Relative humidity Between 20% and 75%
Note: Films may not become wet!
Atmospheric pressure 70 kPa - 106 kPa
Related altitude on site 3000 m to 0 m
Environmental storage conditions
Climate conditions for storage are in accordance with EN60721-3-1-class 1K4.
Room temperature Between -25°C and 55°C (storage)
Relative humidity Between 10% and 100%
Absolute humidity Between 0.1 g/m
3
and 35 g/m
3
Rate of change of temperature 1°C/min
Atmospheric pressure 70 kPa - 106 kPa
Environmental transport conditions
Climate conditions for transport are in accordance with EN60721-3-2-class
2K4.
Temperature Between -40°C and 70°C (transport)
14 | Drystar 5301, Drystar 5302 | Drystar 5301, Drystar 5302