EasyManua.ls Logo

AGFA Drystar 5301 - Page 15

AGFA Drystar 5301
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Relative humidity not combined
with rapid temperature changes
95% at +45°C
Noise emission (method of measurement in accordance with DIN
45635 part 19)
During operation Max. 64 dBA
In stand-by Max. 54 dBA
Total acoustic A-weighted noise power
During operation 62 dB ( = 6.4Bel = 6.4B)
In stand-by 53 dB ( = 5.3Bel = 5.3B)
Print technology
Direct thermal printing
Reliability
Estimated product life
(if regularly serviced and main-
tained according to Agfa instruc-
tions)
> 5 years
Service interventions Max. 2 interventions / 3 years
Earthquake (standard) Meets the CA requirements
Imaging Array - Diagnostic area
8x10” 8” dimensions 10” dimensions
pixels mm pixels mm
2376 188.65 3070 243.76
10x12” 10” dimensions 12” dimensions
pixels mm pixels mm
3070 243.76 3653 290.05
11x14” 11” dimensions 14” dimensions
pixels mm pixels mm
3348 265.83 4358 346.03
14x14” 14” dimensions 14” dimensions
Drystar 5301, Drystar 5302 | Drystar 5301, Drystar 5302 | 15
2831F EN 20210601 1655

Table of Contents

Other manuals for AGFA Drystar 5301

Related product manuals