Mid density 1,2 ± 0,15
High density 2,0 ± 0,2
12. These charts will be used for the daily quality test.
Related Links
Quality Control Charts on page 124
Preventive maintenance schedule on page 106
Performing the daily QC test on page 102
Drystar 5301, Drystar 5302 | Advanced Operation (key-operator mode) | 99
2831F EN 20210601 1655