Using the 2200 TapeStation System
4
Operating Procedure
9 Select the tubes or wells you wish to run by clicking and dragging the
mouse over the sample locations in the user interface.
• Selected wells will change color from white.
• Selected lanes on the ScreenTape device image will change color.
• Lanes which have been run previously will appear gray.
Damage to the 2200 TapeStation instrument and impact on performance
➔ Ensure the lids have been removed from the sample tubes.
For best sizing precision and accuracy, the user should run the appropriate ladder with the
samples.
If 16 samples are to be analyzed in parallel, the user may insert an electronic ladder for each
application in the TapeStation Analysis Software.
Sizing results may not be as accurate using an electronic ladder.
Electronic ladder functionality is not available for D5000, High Sensitivity D5000, or
Genomic DNA assays.
Each ScreenTape device can be used up to 2 weeks after its first use if it has been stored
upright between 2 – 8 °C (36 – 46 °F).
Simply select the samples in the same manner as for an unused ScreenTape device. The
first sample selected will automatically appear in the first available lane.