EasyManuals Logo

Agilent Technologies 55B AA User Manual

Agilent Technologies 55B AA
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #59 background imageLoading...
Page #59 background image
Interface
Agilent 55B AA Spectrometer User’s Guide 59
NOTE
The element can only be selected by loading an appropriate method (user or
cookbook).
This page includes the following instrument-specific fields:
Instrument mode
Defines the mode you want to use, ‘Absorbance’ or ‘Emission’ - default is
Absorbance.
Active lamp
Defines which lamp position to use - (default is position 1).
Active current
Defines the current for the active lamp - default setting is the recommended
current for the lamp of the selected element.
Standby current
Defines the current for the standby lamp - default is 0.0.
D
2
correction
Defines whether D
2
correction is to be used or not - default is ‘No’.
Gas type
Defines the flame type to be used for the analysis - default is the recommended
gas type for the selected element.
Wavelength
Defines the wavelength to be used for the current element - default is the first
listed wavelength.
NOTE
The listed wavelengths are element and instrument-mode specific.
Other λ
Defines the wavelength to be used when ‘Other’ is selected in the Wavelength field.
Range is 180.0900.0 nm.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 55B AA and is the answer not in the manual?

Agilent Technologies 55B AA Specifications

General IconGeneral
BrandAgilent Technologies
Model55B AA
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals