Agilent 55B AA Spectrometer User’s Guide 59
NOTE
The element can only be selected by loading an appropriate method (user or
cookbook).
This page includes the following instrument-specific fields:
Instrument mode
Defines the mode you want to use, ‘Absorbance’ or ‘Emission’ - default is
Absorbance.
Active lamp
Defines which lamp position to use - (default is position 1).
Active current
Defines the current for the active lamp - default setting is the recommended
current for the lamp of the selected element.
Standby current
Defines the current for the standby lamp - default is 0.0.
D
2
correction
Defines whether D
2
correction is to be used or not - default is ‘No’.
Gas type
Defines the flame type to be used for the analysis - default is the recommended
gas type for the selected element.
Wavelength
Defines the wavelength to be used for the current element - default is the first
listed wavelength.
NOTE
The listed wavelengths are element and instrument-mode specific.
Other λ
Defines the wavelength to be used when ‘Other’ is selected in the Wavelength field.
Range is 180.0900.0 nm.