42 Hardware Installation
2 Installation
Verifying High Mass CI System Performance
Special procedure for CI MSD
Refer to the 5975 Series MSD Hardware Manual.
To check high mass performance, the system makes 1 injection of the high
mass checkout sample and analyzes the results.
For high mass CI systems, verify high mass system performance as follows:
1 Load the high mass checkout method, PFHT.M, and resolve any
configuration conflicts.
2 Load the high mass checkout sample into vial 2.
3 When the GC and MSD are ready, select Checkout/High Mass Check from the
Intrument view.
4 Verify that the rms signal-to-noise ratio meets the specifications described
in the 5975 Series MSD Hardware Manual.