Installation 2
Hardware Installation 43
Familiarizing the Customer
Follow the “Customer Familiarization Checklist” on page 53.
Important points about the CI MSD to cover include:
Cleanliness is the most important factor for success in CI, especially
negative mode. This includes air leaks, water, and residual PFDTD
calibrant.
NCI reagent gas setup:
There are no negative reagent ions formed. To set up for NCI mode, tune in
PCI mode, then load NCI tune file to switch parameters, set gas flow to 40%;
adjust if necessary for specific analyte.
NCI operation:
Reagent gas flow, source temperature, and emission current are the three
most important parameters for optimizing NCI.
No PFDTD calibrant ions are formed in PCI with isobutane or ammonia.
Tune with methane, and copy parameters to new file: for example,
PCINH3.U.
The gas purifier can not be used with ammonia.
Must ballast foreline pump daily if running ammonia. Refer to the
maintenance chapter of the 5975 Series MSD Hardware Manual.
For updated information, see the Agilent Technologies Web site at
http://www.agilent.com/chem.

Table of Contents

Other manuals for Agilent Technologies 5975 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 Series and is the answer not in the manual?

Agilent Technologies 5975 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5975 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals