Checklists 3
Hardware Installation 49
Preparing the vacuum system
Plugs removed from the foreline pump
Foreline pump line voltage verified to match available AC power
Line voltage identification sticker installed on MSD
Oil drip pan under the foreline pump
Foreline hose connected to the pump
Pump exhaust vented to a fume hood or the oil trap installed
Foreline pump power cord plugged into MSD
G3397A High Vacuum Gauge Controller installed (required for CI MSDs)
Plumbing the methane line to the CI MSD
Methane/isobutane gas purifier installed
Regulator pressure set to 20 psig (do not exceed 30 psig).
Gas purifier and lines purged
Methane reagent gas lines plumbed to flow module
Connecting the MSD and the GC
GC power turned off
GC/MSD interface heater cable connected
MSD positioned with interface in GC oven
Sideplate thumbscrews loosened completely; analyzer opened
Column installed in the GC/MSD interface
Remote cable connected between the GC and the MSD
Foreline pump and AC supply power cords connected to the MSD.
MSD plugged in
MSD power on (press on the sideplate)
Foreline pump and the front fan operation verified

Table of Contents

Other manuals for Agilent Technologies 5975 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 Series and is the answer not in the manual?

Agilent Technologies 5975 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5975 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals