EasyManua.ls Logo

Agilent Technologies 5975 Series

Agilent Technologies 5975 Series
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 General Maintenance
5975/5977 Series MSD Operating Manual 127
To Clean the EI HES
Materials needed
Abrasive paper (5061-5896)
Alumina abrasive powder (393706201)
Aluminum foil, clean
Cloths, clean (05980-60051)
Cotton swabs (5080-5400)
Glass beakers, 500 mL
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Solvents
Methanol (reagent-grade)
Methylene chloride (reagent-grade)
Acetone (reagent-grade)
Ultrasonic bath
Procedure
1 Disassemble the EI HES. (See “To Disassemble the EI HES” on page 124.)
2 Collect the following parts to be cleaned: (See Figure 38 on page 128.)
Filament mount
Source mount (does not get abrasively or ultrasonically cleaned)
Repeller
Source body
Extractor lens (5)
Post extractor lens 1 (4)
Post extractor lens 2 (3)
Ion focus lens (2)
Entrance lens (1)
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.

Table of Contents

Other manuals for Agilent Technologies 5975 Series

Related product manuals