5 General Maintenance
130 5975/5977 Series MSD Operating Manual
To Assemble the EI HES
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
T6 Torx driver, (8710-2548)
Tweezers (8710-2460)
Procedure
Refer to the exploded parts view Figure 41 and the EI HES parts list Table 17 on
page 133 while using this procedure.
1 Assemble the five lenses inside the lens insulator. (See Figure 39 on
page 131.) The lens number is engraved into the outer circumference of each
lens.
a Starting with the entrance lens 1, set the lens into the end groove in the
lens insulator, and rotate the lens until you feel the ball seat into the
circular recess.
b Insert the next 4 lenses, in numerical order, into the lens insulator. The
lens chamber open end always faces the entrance lens 1. Rotate each
lens until you feel the ball seat into the circular recess.
It is easier to insert lens 5 on an angle because the lens stack at this
point makes the lens insulator less flexible.
CAUTION
Always wear clean gloves when working in the analyzer chamber to avoid
contamination.

Table of Contents

Other manuals for Agilent Technologies 5975 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 Series and is the answer not in the manual?

Agilent Technologies 5975 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5975 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals