EasyManua.ls Logo

Agilent Technologies 5975 Series

Agilent Technologies 5975 Series
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 General Maintenance
5975/5977 Series MSD Operating Manual 147
To Clean an EI XTR, SS, or Inert Source
Materials needed
Abrasive paper (5061-5896)
Alumina abrasive powder (8660-0791)
Aluminum foil, clean
Cloths, clean (05980-60051)
Cotton swabs (5080-5400)
Glass beakers, 500 mL
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Solvents
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
Ultrasonic bath
Preparation
1 Disassemble the ion source. (See “To Disassemble the EI SS or EI Inert
Source” on page 141 or “To Disassemble the EI XTR Source” on page 144.)
2 Collect the following parts to be cleaned for an EI SS or Inert source: (See
Figure 46 on page 149.)
Repeller
Interface socket
Source body
Repeller block insert
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens

Table of Contents

Other manuals for Agilent Technologies 5975 Series

Related product manuals