5 General Maintenance
5975/5977 Series MSD Operating Manual 147
To Clean an EI XTR, SS, or Inert Source
Materials needed
Abrasive paper (5061-5896)
Alumina abrasive powder (8660-0791)
Aluminum foil, clean
Cloths, clean (05980-60051)
Cotton swabs (5080-5400)
Glass beakers, 500 mL
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Solvents
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
Ultrasonic bath
Preparation
1 Disassemble the ion source. (See “To Disassemble the EI SS or EI Inert
Source” on page 141 or “To Disassemble the EI XTR Source” on page 144.)
2 Collect the following parts to be cleaned for an EI SS or Inert source: (See
Figure 46 on page 149.)
Repeller
Interface socket
Source body
Repeller block insert
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens