4 Operating in EI Mode
5975/5977 Series MSD Operating Manual 81
Configuring the MSD Through the Web User
Interface (WUI)
If your GC does not support LVDS communications with an Agilent GC, you can
use the WUI to configure the MSD network settings. Reasons that a GC does not
support configuring a 5977B MSD’s network settings from the GC control panel
include any of the following:
LVDS communication does not exist between the GC and MSD.
The GC is not an Agilent 7890A+ or 7890B model with the correct firmware.
To change the network settings of the MSD
This procedure assumes that the operator has access to a PC located on the
same LAN subnet as the MSD.
1 Open the top hinged cover on the MSD for accessing the analyzer to view the
eModule mini display readout.
2 Press the MSD On/Off switch to start the instrument. When the instrument
has completed its startup initialization it displays the current IP address
information in the mini display readout and cycles through it for about 10
minutes.
3 Copy the IP address, gateway, and subnet mask from the mini display
readout. The defaults for each are 192.168.254.12, 0.0.0.0, and
255.255.255.0 respectively.
4 Enter the IP address into a PC web browser URL to display the WUI page.
(See Figure 21 on page 82.)

Table of Contents

Other manuals for Agilent Technologies 5975 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 Series and is the answer not in the manual?

Agilent Technologies 5975 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5975 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals