2 Installing 7890/7820/6890/6850 GC Columns
50 5975/5977 Series MSD Operating Manual
The GC/MSD interface is heated by an electric cartridge heater. Normally, the
heater is powered and controlled by Thermal Aux #2 heated zone of the GC. The
interface temperature can be set from Agilent MassHunter GC/MS Acquisition
software or from the GC. A thermocouple sensor in the interface monitors the
temperature.
The GC/MSD interface should be operated in the range of 250 to 350 C. The
interface temperature should be slightly higher than the maximum GC oven
temperature, but never higher than the maximum column temperature.
The GC/MSD interface can be used with both EI and CI sources. The EI XTR, HES,
and CI sources require a tip seal assembly. The tip seal and spring interfere with,
and must be removed when using the EI SS and Inert sources.
(See “To Install a Capillary Column in the GC/MS Interface Using the
Self-Tightening Column Nut” on page 36, and “To Install a Capillary Column in
the GC/MS Interface Using a Standard Column Nut” on page 41.)
WARNING
The GC/MSD interface operates at high temperatures. If you touch it when it is
hot, it will burn you.

Table of Contents

Other manuals for Agilent Technologies 5975 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5975 Series and is the answer not in the manual?

Agilent Technologies 5975 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model5975 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals