2 Hardware Overview
Instrument Status Indicators in Data Acquisition
Agilent 6475 Triple Quad LC/MS System User Guide 19
Instrument Status Indicators in Data Acquisition
In MassHunter Data Acquisition, the Instrument Status window indicates the
state of the instrument.
Gray - Not Connected: The instrument is configured with the system and
available to use, but the AIC/workstation is not currently running. The amount of
time for an instrument to reach the Idle state depends on each instrument.
Yellow - Unknown: The device is in an unknown run state.
Dark Yellow - Not Ready: The instrument is connected but is not ready to run
(due to not reaching the correct temperature or pressure required by the method,
for example).
Green - Idle: The instrument is on and ready to process samples.
Teal - Standby: The instrument is in a standby/sleep state.
Purple - Pre-run/Injecting: The instrument is on and is preparing to start
acquisition.
Blue - Running, Post Run: The instrument is currently collecting data.
Orange - Tuning: The instrument is in the process of adjusting MS parameters.
Light Orange - Not Ready: The device is not ready
Red - Instrument Error: The instrument has an error and cannot process
samples.
Blank - Disconnected: The connection to the instrument has been closed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6475 and is the answer not in the manual?

Agilent Technologies 6475 Specifications

General IconGeneral
BrandAgilent Technologies
Model6475
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals