3Tuning
Scheduling a Checktune
Agilent 6475 Triple Quad LC/MS System User Guide 34
Scheduling a Checktune
To specify to run a Checktune automatically,
1 If needed, display the Instrument Status window. On the tool ribbon, click
Instrument Status.
2 Right-click the TQ device in the Instrument Status window.
3 Click Schedule Tune. The Schedule Tune Dialog Box opens.
4 Select Checktune in the left pane. The right pane shows the information for
scheduling a Checktune.
5 Click the Scheduling slider to switch Scheduling on.
6 Select either Weekly or Monthly.
7 Select other options to indicate how often to schedule the Checktune.
8 Click Save.