5 Maintenance
Periodic Tasks
Agilent 6475 Triple Quad LC/MS System User Guide 66
Periodic Tasks
Calibrant Delivery System and Divert Value
The following procedures describe periodic maintenance tasks for the calibrant
delivery system (CDS) and divert values.
Filling a calibrant bottle
Perform this procedure as needed.
Parts required:
APCI calibrant
APCI-L tuning mix
ESI-L tuning mix
MMI-L tuning mix
Reference mix
1 Rinse the calibrant bottle before refilling.
2 Refill the calibrant bottle with the appropriate tuning mix.
3 Put the intake tube into the calibrant bottle as you lift the calibrant bottle into
position.
CAUTION
Do not touch the intake tube unless you are wearing clean gloves.
4 Attach the calibrant bottle onto the fixed bottle cap. Turn the calibrant bottle
counter clockwise to tighten.
CAUTION
The bottle only needs to be snug. Do not over-tighten the bottle. A leaky bottle
cap can prevent the bottle from pressurizing and can lead to the evaporation of
the contents.
CAUTION
The tuning mixes are not interchangeable. Failure to change the tuning mix
when the source or inlet assembly is changed can result in miscalibration of
the instrument and errors in mass assignments.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6475 and is the answer not in the manual?

Agilent Technologies 6475 Specifications

General IconGeneral
BrandAgilent Technologies
Model6475
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals