3Tuning
Generating a Detailed Tune Report
Agilent 6475 Triple Quad LC/MS System User Guide 32
Generating a Detailed Tune Report
A detailed tune report can be generated after you have run Autotune or
Checktune. This is only supported on Ultivo and 6475 LC/TQs.
1 If necessary, open the Method Editor window.
2 Click the TQ tab.
3 Click the Tune > Autotune section in the left pane.
4 Click in the toolbar in the Autotune section. This button locks control of
the TQ instrument. You cannot start a single sample run or a worklist when
Tune has control of the TQ instrument.
5 Click the Generate Detailed Tune Report icon in the toolbar in the TQ
Autotune section. If this button is not available, then you need to run either a
Checktune or an Autotune first.
6 Click in the toolbar to release control of the TQ instrument.