4 Basic Operation
Setting Up and Running an Acquisition Method
Agilent 6475 Triple Quad LC/MS System User Guide 52
Setting Up and Running an Acquisition Method
1 Set up the method:
a Type the values and settings for each of the tabs below.
b Optional. To download the settings to the instrument, click Apply.
c To save the method, click Save As.
d Name the method and click OK.
2 Enter values for all of the LC modules configured for the instrument.
NOTE
Make sure when you type the MS parameters on the next page that the tune file
is the one that you want to use with the acquisition.
3 Enter the TQ parameter values.
a Select the Scan Type from the list in the Time segments table. The Scan
segments table is cleared when you change the Scan Type. The
parameters available on the right change depending on the Scan Type for
the selected Time segment.
If you are changing the Scan Type from MRM to Dynamic MRM or to
Triggered MRM, you can copy and paste the transitions from the original
Scan segments table to the Clipboard and then to the new Scan segments
table. See the online Help for more information.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6475 and is the answer not in the manual?

Agilent Technologies 6475 Specifications

General IconGeneral
BrandAgilent Technologies
Model6475
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals