5 Maintenance
Nebulizer
Agilent 6475 Triple Quad LC/MS System User Guide 73
Nebulizer
The following procedures describe periodic maintenance for the nebulizer.
Removing the nebulizer
Perform this procedure when you need to access the nebulizer for maintenance.
Tools needed:
Clean powder-free nitrile gloves
To remove the nebulizer,
1 Put on clean powder-free nitrile gloves.
2 In the Data Acquisition program, shut off the LC solve flow and nebulizing gas.
3 Remove the left side cover.
4 Disconnect the nebulizer sample line from the nebulizer.
5 Disconnect the sample nebulizer gas tubing from the nebulizer.
WARNING
The tip of the nebulizer can be very hot. Do not touch the nebulizer until it is
cool.
6 Turn the nebulizer counterclockwise until it disengages from the retaining
screws.
7 Gently lift the nebulizer out of the spray chamber.
Inspecting the nebulizer
Perform this procedure if you have the G1960-67470 Nebulizer Adjustment Kit.
Perform this project if there is increased LC back pressure, off-axis spraying,
dripping from the nebulizer, or when the reference nebulizer needle is not
spraying.
Tools needed:
Nebulizer Adjustment Fixture
Clean powder-free nitrile gloves
T6 Torx driver
1/4-inch x 5/16-inch open-end wrench (Quantity: 2)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6475 and is the answer not in the manual?

Agilent Technologies 6475 Specifications

General IconGeneral
BrandAgilent Technologies
Model6475
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals