5 Maintenance
Sources
Agilent 6475 Triple Quad LC/MS System User Guide 62
Sources
The following procedures describe routine maintenance for the components that
make up the source.
Cleaning the spray chamber - daily
Perform this maintenance daily or at the end of each shift or anytime you suspect
carryover contamination from one sample or analysis to another.
NOTE
Use the weekly cleaning procedure if symptoms of contamination persist or if the
spray shield or capillary cap show significant discoloration that cannot be
removed by the regular, daily cleaning.
Tools needed:
Clean powder-free nitrile gloves
Clean lint-free cloth
LC/MS-grade isopropanol
LC/MS-grade water
Clean wash bottle
To perform daily cleaning of the spray chamber,
1 Put on clean powder-free nitrile gloves.
2 Lower the drying gas temperature to the minimum level.
3 For APCI/multimode, lower the vaporizer heater temperature to the minimum
level.
WARNING
Make sure the source is cooled before you clean the spray shield.
4 In a clean wash bottle, prepare a 50:50 mix of LC/MS-grade isopropanol and
LC/MS-grade water.
5 Open the spray chamber.
6 Dampen a clean lint-free cloth with a mixture of LC/MS-grade isopropanol and
LC/MS-grade water.
7 Wipe the spray shield and the area around the spray shield.
8 Close the spray chamber.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6475 and is the answer not in the manual?

Agilent Technologies 6475 Specifications

General IconGeneral
BrandAgilent Technologies
Model6475
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals