EasyManuals Logo

Agilent Technologies 6475 User Manual

Agilent Technologies 6475
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
2 Hardware Overview
Standby mode
Agilent 6475 Triple Quad LC/MS System User Guide 21
Standby mode
Standby mode lets the instrument remain idle for period lasting from minutes to
days. Put the instrument in Standby mode when it is not in use or when you want
to change the ion source.
When the instrument is in Standby mode:
The source interface is ON, the drying gas remains heated, and nebulizer
flows are maintained.
The source and ion optics voltages are turned OFF, and the mass
spectrometer is not sending spectra to the computer.
The LC/MS high voltage electronics are ON.
For APCI and multimode sources: The vaporizer gas heaters are turned down.
For the Agilent Jet Stream (AJS) source: The sheath gas temperature, drying
gas flow, and drying gas temperature are reduced.
The spray chamber high voltages are turned OFF.
The mass spectrometer stop generating spectra.