4 Basic Operation
Turning off the instrument
Agilent 6475 Triple Quad LC/MS System User Guide 40
3 In MassHunter Data Acquisition, in the Tune view or context, right-click the TQ
icon in the Instrument Status window and select Vent to vent the system.
Click Yes to continue.
When you vent the system, the following occurs:
The spray chamber high voltages, the drying gas heater, the nebulization
flow, the detector, and other lens voltages are turned OFF.
The drying gas flow is set to Standby.
If the APCI source is installed, the vaporizer heater is also turned OFF.
If the installed source is an Agilent Jet Stream source, the sheath gas
heater is also turned OFF.
The Rough Pump turns OFF when the turbo speed is below 20 percent.
4 Wait until the instrument completely vents (reaching 760 Torr).
5 Close the Data Acquisition program and shut down the computer.
6 Press the power switch located in the lower left front corner of the instrument.
7 Unplug the main power cable behind the service panel. This action prevents
the turbo pump components from overheating.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6475 and is the answer not in the manual?

Agilent Technologies 6475 Specifications

General IconGeneral
BrandAgilent Technologies
Model6475
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals