5 Maintenance
Rough Pump
Agilent 6475 Triple Quad LC/MS System User Guide 88
Replacing fuses or resetting circuit breakers
Perform this procedure as needed for maintenance.
Tools needed:
Long flat-blade screwdriver
8 Amp fuse
12.5 Amp fuse
WARNING
Never replace a fuse while the instrument is plugged into the power outlet. You
run the risk of an electric shock.
CAUTION
If the instrument is plugged in and turned on when you change a fuse, you will
damage the instrument.
To replace fuses or reset circuit breakers,
1 Unplug the instrument power cord or cords from the power outlet.
2 If your LC/MS instrument includes circuit breakers and you need to reset
them, press in the circuit breaker buttons.
3 Using a flat-blade screw driver, remove the fuse holder of the blown fuse.
4 Replace with the appropriate fuse.
5 Reinstall the fuse holder.
6 Plug in the instrument.