EasyManua.ls Logo

Agilent Technologies 7890A - To Verify FID Ignitor Function During Ignition Sequence

Agilent Technologies 7890A
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting Tasks 8
Troubleshooting 105
To Verify FID Ignitor Function During Ignition Sequence
1 Remove the detector top cover.
2 Turn the FID flame On.
3 Observe the ignitor plug though the FID chimney. The
small hole should glow during ignition sequence.
WARNING
Keep body parts at a safe distance from the FID chimney while
performing this task. If using hydrogen, the FID flame will not be
visible.

Table of Contents

Other manuals for Agilent Technologies 7890A

Related product manuals