EasyManua.ls Logo

Agilent Technologies 7890A - The Measured Column Flow Does Not Equal the Displayed Flow

Agilent Technologies 7890A
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
70 Troubleshooting
4 GC Not Ready Symptoms
The Measured Column Flow Does Not Equal the Displayed Flow
If the actual column flow does not match the calculated flow
displayed on the GC, do the following:
Verify that the measured flows are corrected to 25 °C and
1 atmosphere.
Verify that the correct column dimensions are configured
accurately, including the actual (trimmed) column length.
The split vent line or trap may be partly plugged, creating
an actual inlet pressure higher than the setpoint pressure.

Table of Contents

Other manuals for Agilent Technologies 7890A

Related product manuals