Chromatographic Symptoms 3
Troubleshooting 37
Peak Fronting
The figure below shows examples of the three types of
peaks: symmetric, fronting, and overloaded.
If peak fronting or overloading occurs, try the following:
Verify that the injection volume is appropriate.
Ensure that the column is installed properly.
Verify that the appropriate injection technique is being
used.
If using capillary splitless injection, consider the
compound solubility in the injection solvent.
Change the solvent.
Use a retention gap.
Check purity of sample solvent.
Symmetrical Fronting Overloaded