36 Troubleshooting
3 Chromatographic Symptoms
Sample Decomposition in Inlet/Missing Peaks
Lower the inlet temperature.
Check for air or water in the carrier gas; verify gas purity
and functionality of traps.
Verify that the liner is appropriate for the sample being
run.
Perform complete inlet maintenance: Replace all
consumable parts and bake out the inlet.
Install a deactivated liner (SS, PP, MMI and PTV inlets).
Check for leaks at the septum, liner, and column fittings.
(See “Checking for Leaks”.)
Install an Agilent Direct Connect liner.
Use a pulsed pressure method for quicker sample transfer
to column.
Bake out the inlet. See the following:
To Bakeout Contaminants from the Split/Splitless Inlet
To Bakeout Contaminants from the Multimode Inlet
To Bakeout Contaminants from the Purged Packed Inlet
To Bakeout Contaminants from the COC Inlet
To Bakeout Contaminants from the PTV Inlet
To Bakeout Contaminants from the VI Inlet