A Gas Exceeds Pressure Setpoint or Flow
If a gas exceeds its pressure or flow setpoint, do the
following:
If using a split/splitless inlet, MMI inlet, PTV inlet, or
volatiles interface:
Decrease the split ratio.
Replace the split vent filter.
Verify that the correct liner is selected (for split/splitless,
MMI and PTV inlets).
Check the gold seal for contamination (for split/splitless
inlet).
If using an FID or NPD:
Check for a plugged jet.
Valves:
Check for a misaligned rotor.