Chromatographic Symptoms 3
Troubleshooting 49
FPD Baseline Output High, > 20 pA
Check the supply gas purity.
Check all trap indicators and dates.
Check the detector for contamination.
Check for light leaks at the photomultiplier tube (PMT);
tighten the PMT if it is loose.
Perform complete inlet maintenance: Replace all
consumable parts and bake out the inlet.
Perform column maintenance: Bake out contaminants as
needed.