EasyManua.ls Logo

Agilent Technologies 7890A - FPD Flame Goes Out During a Run and Attempts to Reignite

Agilent Technologies 7890A
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chromatographic Symptoms 3
Troubleshooting 45
FPD Flame Goes Out During a Run and Attempts to Reignite
If the flame goes out during a run, do the following:
Check the GC system for leaks, especially at the detector
column fitting. (See “Checking for Leaks”.)
Verify the detector temperature is set
200 °C.
Ensure that condensation in the vent tube is not dripping
back into the detector. The flexible plastic vent tube must
run from the detector into a container, without sagging,
in order to properly drain water condensate. Keep the
open tube end out of the water in the container.
If the FPD flame goes out and then reignites, do the
following:
Verify that the Lit offset setting is lower than the normal
baseline.
Check for leaks. (See “Checking for Leaks”.)
Check the flow settings, then measure the actual detector
flows. (See “To Measure a Detector Flow”.)

Table of Contents

Other manuals for Agilent Technologies 7890A

Related product manuals