EasyManua.ls Logo

Agilent Technologies 7890A - FID Flame Goes Out During a Run and Attempts to Reignite

Agilent Technologies 7890A
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chromatographic Symptoms 3
Troubleshooting 43
FID Flame Goes Out During a Run and Attempts to Reignite
The following is an example chromatogram showing a
flameout from a large solvent peak.
After a flameout, the GC will try to ignite the flame three
times.
If the FID flame goes out during a run, do the following:
See if an aromatic peak or water extinguished the flame.
Check for a plugged jet.
Verify that the gas flow settings are correct. Verify that Lit
offset is set appropriately.
If the FID flame attempts to reignite but is already lit, do
the following:
Verify that the FID Lit offset setting is appropriate for the
run (typically
< 2.0 pA).
Check to see if an aromatic peak or water extinguished
the flame.
Check for a partially plugged jet. Measure actual
hydrogen, air, and makeup flows at the detector. (See “To
Measure a Detector Flow”.)
Check for leaks at the detector column fitting. (See
“Checking for Leaks”.)
9
8
7
6
5
4
3
2
1
23
4
5min
pA

Table of Contents

Other manuals for Agilent Technologies 7890A

Related product manuals