To Check for External Leaks
Check for leaks at these connections:
Gas supply bulkhead fittings
Gas cylinder fitting
Regulator fittings
Traps
Shut- off valves
T- fittings
Perform a pressure drop test.
1 Turn off the GC.
2 Set the regulator pressure to 415 kPa (60 psi).