44 Troubleshooting
3 Chromatographic Symptoms
FID Baseline Output Above 20 pA
Verify the purity of the carrier and detector gas supply.
Inspect the column for column bleed.
Check the gas supply trap indicators/dates and ensure
that the traps are not expended.
Verify that the detector was reassembled properly after
recent maintenance.
Inspect the detector for contamination.
Check that the FID leakage current is < 2.0 pA. (See “To
Measure FID Leakage Current”.)