68 Troubleshooting
4 GC Not Ready Symptoms
The Inlet Pressure or Flow Fluctuates
A fluctuation in inlet pressure causes variations in the flow
rate and retention times during a run. Do the following:
Check if the gas purifier or gas generator is operating at
or near capacity.
Check the supply gas for sufficient delivery pressure.
Verify that the supply pressure regulator is functioning
properly.
Check for leaks. (See “Checking for Leaks”.)
Check for large restrictions in the inlet liner or split vent
trap.
Verify that the correct liner is installed.
Check for a restriction in headspace, purge and trap, and
any other external sampling devices.