Troubleshooting Tasks 8
Troubleshooting 111
To Adjust the FPD Lit Offset
To adjust the FPD Lit offset:
1 Press [Config].
2 Scroll to Front detector or Back detector (wherever the
detector is installed) and press [Enter].
3 Scroll to Lit offset. With the Lit offset line highlighted, enter
the new parameter for the FPD (typical value is 2.0 pA),
and press [Enter].
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.