Chromatographic Symptoms 3
Troubleshooting 41
Low Peak Area or Height (Low Sensitivity)
If using an inlet in split mode, check the split ratio.
Check for leaks. (See “Checking for Leaks”.)
Check the inlet for contamination. (See “Contamination or
Carryover”.)
Check each column and verify that it was cut and
installed properly at each end.
Verify that the column type is correct.
Perform column maintenance: Bake out contaminants,
remove the contaminated length of column near the inlet,
and reverse and bake out the column as needed.
Verify that the liner type is appropriate for the sample.
Verify that the detector flow settings are correct.
Check the supply gas purity.
Check all trap indicators and dates.
Verify that the method parameters are correct.
Check sample stability.
Check configured syringe size. Some syringe sizes are
specified at half- capacity. If the maximum syringe volume
is marked at half- height on the barrel, not at the top of
the barrel, enter twice the labeled volume when
configuring the syringe size.
If using an FID:
Verify that the correct jet is installed.
Check for a dirty jet.
If using a uECD:
Replace the fused silica indented mixing liner.
Replace and reinstall column.
Clean the makeup gas adapter.
If using an NPD:
Check the detector for contamination.
Replace ceramic insulators.
Replace the bead.
If using an FPD:
Verify correct column installation.