Agilent Technologies 7890A - Cannot Maintain a Pressure as Low as the Setpoint on a Split Inlet

Agilent Technologies 7890A
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GC Not Ready Symptoms 4
Troubleshooting 69
Cannot Maintain a Pressure as Low as the Setpoint on a Split Inlet
If the GC cannot maintain a pressure as low as the setpoint,
check for the following:
Consider using a liner designed for split analysis.
Check for a plugged liner.
Check for contamination in the split vent line. Contact
Agilent service to replace, if necessary.
For the split/splitless inlet, replace gold seal.