EasyManua.ls Logo

Agilent Technologies 7890A - Column Shutdowns

Agilent Technologies 7890A
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
76 Troubleshooting
5 Shutdown Symptoms
Column Shutdowns
The GC monitors inlet and auxiliary gas streams. If a carrier
gas (which can include an auxiliary flow module or
pneumatics control module) is unable to reach its flow or
pressure setpoint, the GC assumes that a leak exists. It will
warn you with a beep after 25 seconds, and it will continue
to beep in intervals. After about 5 minutes, the GC will shut
down components to create a safe state. The GC:
Displays Front inlet pressure shutdown.
Turns off to avoid column damage.
Opens oven flaps in the back of the oven halfway.
Flashes oven temperature setpoint Off.
Turns off all flows for the column. When viewed, their
parameters flash Off. For example, the septum purge and
column flows for a split/splitless inlet would turn off.
Turns off all other heaters. When viewed, their
temperature parameters flash Off.
Attempts to turn on a shut-down zone fail with an error
message.
To recover from this state.
1 Fix the cause of the shutdown.
Check for a broken column near the inlet.
Check for leaks.
Replace the inlet septum.
Replace the inlet O- ring.
Check the supply pressure.
2 Press the key for the device that initiated the shutdown.
Scroll to the pneumatic parameter that is flashing Off,
then press [On] or [Off].
For example, if the front inlet ran out of carrier gas,
press [Front Inlet], scroll to the pressure or flow parameter,
then press [On].

Table of Contents

Other manuals for Agilent Technologies 7890A

Related product manuals