EasyManua.ls Logo

Agilent Technologies 7890A - FID Ignitor Does Not Glow During Injection Sequence

Agilent Technologies 7890A
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
72 Troubleshooting
4 GC Not Ready Symptoms
FID Ignitor Does Not Glow During Injection Sequence
1 Remove the detector top cover.
2 Turn the FID flame On.
3 Observe the ignitor plug though the FID chimney. The
small hole should glow during ignition sequence.
If the test fails, check for the following possible causes:
The ignitor may be bad; replace the ignitor.
Detector temperature is set to < 150 °C. Agilent
recommends operating the FID at >
300 °C.
The ignitor is not making a good connection to the
ground:
The ignitor must be tightly screwed into the FID castle
assembly.
The three T- 20 Torx screws that hold the collector
assembly in place must be tight.
The brass knurled nut that holds the FID castle
assembly in place must be tight.
Perform FID maintenance if these parts are corroded or
oxidized.
WARNING
Keep body parts at a safe distance from the FID chimney while
performing this task. If using hydrogen, the FID flame will not be
visible.
T-20 Torx screws (3)
Ignitor
Knurled nut

Table of Contents

Other manuals for Agilent Technologies 7890A

Related product manuals