FPD Does Not Ignite
Check that the FPD temperature is high enough for
ignition (> 150 °C).
Check FPD flow rates and that they match the type of
filter installed in the FPD.
Measure the actual detector flows. (See To Measure a
Detector Flow.)
The column may be installed too high into the detector.
Check that the FPD ignitor operates. (See To Verify That
the FID Flame Is Lit.)
Check the column and makeup flow rates.
Ensure that condensation in the vent tube is not dripping
back into the detector. The flexible plastic vent tube must
run from the detector into a container, without sagging,
in order to properly drain water condensate. Keep the
open tube end out of the water in the container.
Check the Lit offset value. The typical Lit offset value is 2.0.
Thoroughly leak check the whole system. (See Checking
for Leaks.)