30 Troubleshooting
3 Chromatographic Symptoms
Larger Peaks Than Expected
Check each configured column’s dimensions against the
actual column dimensions. (See “Configurable Items to
Always Keep Current”.)
Check the autosampler injection volume.
Check the vial caps.
Check configured syringe size. Some syringe sizes are
specified at half- capacity. If the maximum syringe volume
is marked at half- height on the barrel, not at the top of
the barrel, enter twice the labeled volume when
configuring the syringe size.