Troubleshooting Tasks 8
Troubleshooting 99
For the FPD, remove the plastic tubing from the FPD
exhaust and connect the flowmeter directly to the FPD
vent tube. If necessary, use a 1/4- inch tube adapter
between the detector exhaust and the flowmeter tubing.
7 Connect the flowmeter to the flowmeter adapter.
8 Measure the actual flow rate of each gas one at a time.