EasyManua.ls Logo

Agilent Technologies 7890A - To Start up the GC

Agilent Technologies 7890A
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Basics 2
Operating Guide 27
To Start Up the GC
Successful operation begins with a properly installed and
maintained GC. The utility requirements for gases, power
supply, venting of hazardous chemicals, and required
operational clearances around the GC are detailed in the
Site Preparation Checklist.
1 Check gas source pressures. For required pressures, see
the Agilent GC, MSD, and ALS Site Preparation Checklist.
2 Turn on the carrier and detector gases at their sources
and open the local shutoff valves.
3 Turn on the cryo coolant at its source, if used.
4 Turn on the GC power. Wait for Power on successful to be
displayed.
5 If removed, install the column.
6 Check that the column fittings are leak free.
7 Load the analytical method.
8 Wait for the detector(s) to stabilize before acquiring data.
The time required for the detector to reach a stable
condition depends on whether the detector was turned off
or its temperature was reduced while the detector
remained powered.
Detector type Stabilization time starting
from a reduced temperature
(hours)
Stabilization time starting
from detector off
(hours)
FID 2 4
TCD 2 4
ECD 4 18 to 24
FPD 2 12
NPD 4 18 to 24

Table of Contents

Other manuals for Agilent Technologies 7890A

Related product manuals