34 Operating Guide
3 Running a Method or a Sequence from the Keypad
Loading, Storing, and Running Sequences from the Keypad
A sequence can specify up to five subsequences to be run,
as well as priority (ALS only) and post- run sequences, if
defined. Each sequence is stored as a number (from 1 to 9).
To load a sequence
1 Press [Load][Seq].
2 Enter the number of the sequence to be loaded.
3 Press [On/Yes] to load the sequence or [Off/No] to cancel
the load.
An error message is displayed if the specified sequence
number has not been stored.
To store a sequence
1 Press [Store][Seq].
2 Enter the number of the sequence to be stored.
3 Press [On/Yes] to store the sequence or [Off/No] to cancel
the store.
To start running a sequence
1 Load the sequence. (See "To load a sequence".)
2 Press [Seq Control].
3 Verify the status of the sequence:
Runningthe sequence is running
Ready/waitthe instrument is not ready (due to oven
temperature, equilibration times, etc.)
Pausedthe sequence is paused
Stoppedproceed to step 4
Abortedthe sequence stopped without waiting for the
run to finish (See "Aborting a sequence".)
No sequencethe sequence is off or not defined
4 Scroll to the Start sequence line and press [Enter] to change
the status to Running.