Operating the 7693A Automatic Sampler 5
Operating Guide 63
To Adapt the 7693A ALS for the COC Inlet
The 7693A injectors can inject samples directly into 250- μm,
320-
μm, and 530- μm columns in GCs with a cool- on- column
inlet.
When performing cool on- column injections, the injector:
Slows the carriage speed so the overall injection time
increases to 500 milliseconds.
Lowers the tip of the syringe needle an additional 19 mm
into the column.
To adapt the injector and GC for cool on- column use, follow
these steps:
1 If necessary, remove the current syringe from the injector.
See “To Remove the Syringe on the 7693A ALS” for more
information.
2 If necessary, remove the injector from the mounting post
and unplug the injector cable. If desired, park it on a
parking post.
3 Select the on- column syringe needed for the column size.
See your Agilent catalog for consumables and supplies,
your GC operating documentation, and the Agilent web
site (www.agilent.com/chem) for a list of parts.
4 Replace the needle support foot with the supplied
on- column foot. See the Agilent 7693A Automatic Liquid
Sampler Installation, Operation, and Maintenance manual
for more information.
5 Install the on- column syringe. See “To Install a Syringe on
the 7693A ALS” for more information.
6 Prepare the GC inlet. See your GC operating
documentation for instructions.
Check your needle to column size.
Verify that the insert matches the needle size.
If necessary, replace the septum.
For a list of other consumables needed to perform these
injections, see your GC operating documentation.
7 Plug in the injector cable.
8 Install the injector on the mounting post.