32 Operating Guide
3 Running a Method or a Sequence from the Keypad
Loading, Storing, and Running Methods from the Keypad
To load a method
1 Press [Load].
2 Press [Method].
3 Enter the number of the method to be loaded (1 through
9).
4 Press [On/Yes] to load the method and replace the active
method. Alternatively, press [Off/No] to return to the
stored methods list without loading the method.
To store a method
1 Ensure that the proper parameters are set.
2 Press [Store].
3 Press [Method].
4 Enter the number of the method to store (1 through 9).
5 Press [On/Yes] to store the method and replace the active
method. Alternatively, press [Off/No] to return to the
stored methods list without storing the method.
To manually inject a sample with a syringe and start a run
1 Prepare the sample syringe for injection.
2 Load the desired method. (See "To load a method".)
3 Press [Prep Run].
4 Wait for STATUS Ready for Injection to be displayed.
5 Insert the syringe needle into the inlet and puncture the
septum.
6 Simultaneously depress the syringe plunger to inject the
sample and press [Start].
To run a method to process a single ALS sample
1 Prepare the sample for injection.
2 Load the sample vial into the assigned location in the
ALS tray or turret.
3 Load the desired method. (See "To load a method".)
4 Press [Start] on the GC keypad to initiate the ALS syringe
cleaning, sample loading, and sample injection method.