ACPALTCH Adjacent Channel Power Alternate Channels
ACPALTCH Adjacent Channel Power
Alternate Channels
Syntax
Figure 7-8 ACPALTCH Syntax
Description
The ACPALTCH command sets the number of alternate channels to be
measured by an adjacent channel power measurement to either 0, 1, or
2. Zero is the default. The number of alternate channels is used with
the ACPRSLTS command.
Query Response
Figure 7-9 ACPALTCH Query Response
Example
10 INTEGER Altchannel
20 Altchannel = 2
30 OUTPUT 718;"ACPALTCH";Altchannel;";"
40 END