EasyManua.ls Logo

Agilent Technologies 8560E - MKSP Marker Delta to Span; Figure 7-183 MKSP Syntax

Agilent Technologies 8560E
716 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
536 Chapter7
Language Reference
MKSP Marker Delta to Span
MKSP Marker Delta to Span
Syntax
Figure 7-183 MKSP Syntax
Description
The MKSP command sets the frequency span equal to the frequency
difference between two markers on a trace. The start frequency is set
equal to the frequency of the left-most marker and the stop frequency is
set equal to the frequency of the right-most marker.
Example
10 INPUT "CONNECT THE 300 MHZ CALIBRATOR TO THE INPUT",Ans
$
20 OUTPUT 718;"IP;SNGLS;FA 270MHZ;FB 1275MHZ;TS;"
30 OUTPUT 718;"MKPK HI;MKD;MKPK NH;TS;MKSP;TS;"
40 END